40 Watt Medical Power Supplies

Product Information

MEDIMOD® AJ-040UM SERIES is a 40 Watt Medical Power Supplies platform with both standard and configurable models featuring output voltage(s) that can be quickly configured to order while maintaining all international safety approvals.

These 40 watt medical power supplies are available in single, dual or triple output configurations with outputs ranging from 1.5 to 48 VDC. The AJ-040UM features an industry-standard footprint; universal AC input; international safety approvals; Class B emissions; and -20°C to +70°C operation.

MEDIMOD® 40 watt medical power supplies capabilities include a multitude of output voltage configurations, extended temperature operating range, attached wire harnesses and much, much more. All these modifications are available without any impact on safety approvals to reduce both development cost and time to market.

unipower-gold-seal
FIVE YEAR WARRANTY
Features and Options
  • Universal AC Input Range
  • High-Density 3.6” x 2.5” Footprint
  • 1 or 3 Outputs configurable from 1.5~48VDC
  • Five Year Warranty
  • Medical Safety Approvals
  • >500k Hours MTBF Demonstrated
  • Optional -40°C Guaranteed Start-Up
  • Double sided PC board
SPECIFICATIONS
 
INPUT
Voltage Range 90-264VAC
Frequency 47-63Hz
Inrush Current 25A Max (cold start)
Current @ 115VAC 1A max
Leakage Current <100μA @ 264VAC
Fusing2A / 250VAC
 
OUTPUT
Power

30W Convection | 40W with 15 cfm Airflow

Hold-up Time16mSec
Efficiency70% Typical
Adjustment Range (V1 Only)±5%
Ripple / Noise, max. 1% pk-pk max.
Line Regulation ±0.2% max.
Load Regulation @ 60% ±40% Full Load

V1: ±3% max.
V2-V3: ±5% max.

Cross Regulation @ 60% ± 40% Full Load V1: Change in V2 - V3: ±0.5%
V2 - V3: Change in V1 @75 ±25% F/L: ±5% max.
Overvoltage Protection (V1 Only) >125% (Latch Off)
Power Limit >120% (Auto-Recovery)
Overshoot (all outputs) 10% max.
Response Time 500μSec (25-75% step load)
Switching Frequency 60KHz (typical)
 
ENVIRONMENTAL
Operating Temp. Range-20°C to +50°C (Full Load)
Consult factory for -40°C Guaranteed Start-Up and Industrial Temperature Range options
Output Current Derating2.5%/°C, 50°C to 70°C
Storage Temp. Range -40°C to + 85°C
Humidity 5% to 95%, Non-Condensing
MTBF, Demonstrated >500,000 Hours
Cooling 15 cfm Airflow for Full Power
ImmunityEN61000-4-2; -3; -4; -5; -6; -8; -11
Altitude10,000 feet
 
PHYSICAL SPECIFICATIONS
Dimensions (see datasheet for details) 3.60 x 2.44 x 1.00” / 91.5 x 62.0 x 25.4mm
Weight0.38 lbs. (0.17 kg.)
Vibration from 10 - 55Hz1.0G Peak
.(3 orthogonal axes @ 1 octave/min, 5 minute dwell @ 4 major resonances)
 
COMPLIANCE
EMCFCC Class A & VDE Class A, CISPR 11; EN 55011 Class A
SafetyUL60601-1 3rd Ed., EN60601-1 3rd Ed., CB REPORT (IEC 60601-1), CE MARK (LVD)

“IF WHAT YOU SEE IS WHAT YOU DON’T WANT, IT CAN EASILY BE CHANGED.”

The MEDIMOD® family of switched mode medical power supplies has been designed with two precepts; (1) the laws of physics are immutable, and (2) the satisfaction of customer requirements and needs is paramount.

A host of modifications, only some of which are listed below, can and will be performed on products for customer programs requiring as few as 250 units per year. These “mods” are available at nominal premium (if any), normally without non-recurring engineering costs (although a one time documentation fee may be incurred), and usually with all safety agency approvals in place. This minimizes both product development cost and new product time to market. Effectively, MEDIMODs® medical switchers allow small program requirements the luxury of costly custom power supply designs.

TYPICAL MODIFICATIONS

  • Unique Output Combinations from 1.5 to >48 volts
  • Low Output Ripple and Noise
  • Extended Temperature Operating Range
  • -40°C Start-Up
  • Zero Load Operation